Evaluation of the i-STAT analyzer for determination of ionized calcium concentrations in bovine blood
| dc.contributor.author | Yilmaz, Oznur | |
| dc.contributor.author | Karapinar, Tolga | |
| dc.date.accessioned | 2026-08-12T17:17:52Z | |
| dc.date.issued | 2019 | |
| dc.department | Fırat Üniversitesi | |
| dc.description.abstract | Background Information regarding the performance of the i-STAT, a point-of-care analyzer, to determine ionized calcium (iCa(2+)) concentration in bovine blood is limited. Objective The primary goal of this study was to evaluate the agreement of the i-STAT against a traditional bench-top blood gas analyzer (Radiometer ABL800 FLEX) in the measurement of iCa(2+) in the blood of cattle with various disorders. Methods In this study, 121 cattle with various diseases were used. Venous blood samples were run on the i-STAT and the Radiometer ABL800 FLEX for determination of blood iCa(2+) concentrations. Data were subjected to Passing-Bablok regression and Bland-Altman plots. The Radiometer ABL800 FLEX was the reference method and the i-STAT was the test method. The i-STAT performance was evaluated at two different cutoff values for iCa(2+) concentrations (<1.00 and <1.18 mmol/L) using the receiver operating characteristic curve. Results The Passing-Bablok regression equation that fit the iCa(2+) concentration as measured with the i-STAT (Y) and the Radiometer ABL800 FLEX (X), was Y = -0.108276 [-0.15/-0.06] + 1.103448 [1.06/1.14] x X, with a residual deviation of 0.02. The deviation from the identity line was not significant (P = 0.64). The Bland-Altman difference plot revealed no bias between the i-STAT and the Radiometer ABL 800 FLEX. The i-STAT had a sensitivity of 100% (95% CI, 87.7-100.0) and 94.8% (95% CI, 87.2-98.6) at the iCa(2+) concentration cutoff values of <1.00 and <1.18 mmol/L, respectively. Conclusions iCa(2+) concentrations in bovine blood measured with the i-STAT agreed with those of the Radiometer ABL800 FLEX. In particular, the i-STAT predicted hypocalcemia with a high sensitivity. | |
| dc.description.sponsorship | Firat University Scientific Research Unity (FUBAP) [VF.16.24] | |
| dc.description.sponsorship | Firat University Scientific Research Unity (FUBAP), Grant/Award Number: VF.16.24 | |
| dc.identifier.doi | 10.1111/vcp.12705 | |
| dc.identifier.endpage | 35 | |
| dc.identifier.issn | 0275-6382 | |
| dc.identifier.issn | 1939-165X | |
| dc.identifier.issue | 1 | |
| dc.identifier.orcid | 0000-0003-1724-491X | |
| dc.identifier.pmid | 30786050 | |
| dc.identifier.scopus | 2-s2.0-85061931140 | |
| dc.identifier.scopusquality | Q2 | |
| dc.identifier.startpage | 31 | |
| dc.identifier.uri | https://doi.org/10.1111/vcp.12705 | |
| dc.identifier.uri | https://hdl.handle.net/11508/52831 | |
| dc.identifier.volume | 48 | |
| dc.identifier.wos | WOS:000462771600006 | |
| dc.identifier.wosquality | Q3 | |
| dc.indekslendigikaynak | Web of Science | |
| dc.indekslendigikaynak | Scopus | |
| dc.indekslendigikaynak | PubMed | |
| dc.language.iso | en | |
| dc.publisher | Wiley | |
| dc.relation.ispartof | Veterinary Clinical Pathology | |
| dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | |
| dc.rights | info:eu-repo/semantics/closedAccess | |
| dc.snmz | KA_WoS_20260511 | |
| dc.subject | calcium | |
| dc.subject | cattle | |
| dc.subject | cow | |
| dc.subject | point-of-care | |
| dc.subject | sensitivity | |
| dc.title | Evaluation of the i-STAT analyzer for determination of ionized calcium concentrations in bovine blood | |
| dc.type | Article |