Evaluation of the i-STAT analyzer for determination of ionized calcium concentrations in bovine blood

dc.contributor.authorYilmaz, Oznur
dc.contributor.authorKarapinar, Tolga
dc.date.accessioned2026-08-12T17:17:52Z
dc.date.issued2019
dc.departmentFırat Üniversitesi
dc.description.abstractBackground Information regarding the performance of the i-STAT, a point-of-care analyzer, to determine ionized calcium (iCa(2+)) concentration in bovine blood is limited. Objective The primary goal of this study was to evaluate the agreement of the i-STAT against a traditional bench-top blood gas analyzer (Radiometer ABL800 FLEX) in the measurement of iCa(2+) in the blood of cattle with various disorders. Methods In this study, 121 cattle with various diseases were used. Venous blood samples were run on the i-STAT and the Radiometer ABL800 FLEX for determination of blood iCa(2+) concentrations. Data were subjected to Passing-Bablok regression and Bland-Altman plots. The Radiometer ABL800 FLEX was the reference method and the i-STAT was the test method. The i-STAT performance was evaluated at two different cutoff values for iCa(2+) concentrations (<1.00 and <1.18 mmol/L) using the receiver operating characteristic curve. Results The Passing-Bablok regression equation that fit the iCa(2+) concentration as measured with the i-STAT (Y) and the Radiometer ABL800 FLEX (X), was Y = -0.108276 [-0.15/-0.06] + 1.103448 [1.06/1.14] x X, with a residual deviation of 0.02. The deviation from the identity line was not significant (P = 0.64). The Bland-Altman difference plot revealed no bias between the i-STAT and the Radiometer ABL 800 FLEX. The i-STAT had a sensitivity of 100% (95% CI, 87.7-100.0) and 94.8% (95% CI, 87.2-98.6) at the iCa(2+) concentration cutoff values of <1.00 and <1.18 mmol/L, respectively. Conclusions iCa(2+) concentrations in bovine blood measured with the i-STAT agreed with those of the Radiometer ABL800 FLEX. In particular, the i-STAT predicted hypocalcemia with a high sensitivity.
dc.description.sponsorshipFirat University Scientific Research Unity (FUBAP) [VF.16.24]
dc.description.sponsorshipFirat University Scientific Research Unity (FUBAP), Grant/Award Number: VF.16.24
dc.identifier.doi10.1111/vcp.12705
dc.identifier.endpage35
dc.identifier.issn0275-6382
dc.identifier.issn1939-165X
dc.identifier.issue1
dc.identifier.orcid0000-0003-1724-491X
dc.identifier.pmid30786050
dc.identifier.scopus2-s2.0-85061931140
dc.identifier.scopusqualityQ2
dc.identifier.startpage31
dc.identifier.urihttps://doi.org/10.1111/vcp.12705
dc.identifier.urihttps://hdl.handle.net/11508/52831
dc.identifier.volume48
dc.identifier.wosWOS:000462771600006
dc.identifier.wosqualityQ3
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherWiley
dc.relation.ispartofVeterinary Clinical Pathology
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.snmzKA_WoS_20260511
dc.subjectcalcium
dc.subjectcattle
dc.subjectcow
dc.subjectpoint-of-care
dc.subjectsensitivity
dc.titleEvaluation of the i-STAT analyzer for determination of ionized calcium concentrations in bovine blood
dc.typeArticle

Dosyalar