Preliminary results from CdTe spectrum detectors

dc.contributor.authorDo?ru, Mahmut
dc.date.accessioned2026-08-12T16:11:36Z
dc.date.issued1997
dc.departmentFırat Üniversitesi
dc.description.abstractA few CdTe detectors of different thicknesses and active areas have been studied for gamma-ray spectroscopy. The current of the detectors is low even at high bias voltages. The charge collection efficiency (cce) up to 98 % for the detection of 60 keV gamma radiation from 241Am and up to 85 % for the detection of 122 keV gamma radiation from 57Co sources have been obtained. The measured cce is good agreement compared with previously calculated cce using a model which includes trapping effects. The detector radiation count rate increases dramatically with applied bias. Polarization effect is produced by radiating the detectors for extended time with 57Co source. Decreases in count rate is very slow. © Tübi?tak.
dc.identifier.endpage1057
dc.identifier.issn1300-0101
dc.identifier.issue10
dc.identifier.scopus2-s2.0-0006195664
dc.identifier.scopusqualityQ2
dc.identifier.startpage1049
dc.identifier.urihttps://hdl.handle.net/11508/42550
dc.identifier.volume21
dc.indekslendigikaynakScopus
dc.language.isoen
dc.relation.ispartofTurkish Journal of Physics
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.snmzKA_Scopus_20260511
dc.titlePreliminary results from CdTe spectrum detectors
dc.typeArticle

Dosyalar