Metallurgical failure analysis of various implant materials used in orthopedic applications

dc.contributor.authorAksakal, B.
dc.contributor.authorYildirim, Ö.S.
dc.contributor.authorGul, H.
dc.date.accessioned2026-08-12T16:13:43Z
dc.date.issued2004
dc.departmentFırat Üniversitesi
dc.description.abstractImplants made of titanium alloy Ti6Al4V and 316L steel obtained from revision operations between 1993 and 2002 were examined. Focus was on failed implant materials removed from patients undergoing coxarthrosis and scoliosis revision operations. The failed materials were examined metallurgically through a scanning electron microscope (SEM). The type of implant materials, their survival time in the body, their chemical composition, and the reasons for failure were accounted for. Analysis of the results revealed that the main reasons for failure are corrosive attack, manufacturing defects, and nonstandard-caused failures.
dc.identifier.doi10.1361/15477020419794
dc.identifier.endpage23
dc.identifier.issn1547-7029
dc.identifier.issue3
dc.identifier.scopus2-s2.0-12344255851
dc.identifier.scopusqualityQ2
dc.identifier.startpage17
dc.identifier.urihttps://doi.org/10.1361/15477020419794
dc.identifier.urihttps://hdl.handle.net/11508/43192
dc.identifier.volume4
dc.indekslendigikaynakScopus
dc.language.isoen
dc.relation.ispartofJournal of Failure Analysis and Prevention
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.snmzKA_Scopus_20260511
dc.subjectComposition; Corrosion; Defects; Fatigue of materials; Implants (surgical); Metallurgy; Orthopedics; Scanning electron microscopy; Steel; Strength of materials; Titanium alloys; Corrosive attack; Coxarthrosis; Metallurgical failure analysis; Scoliosis; Failure analysis
dc.titleMetallurgical failure analysis of various implant materials used in orthopedic applications
dc.typeArticle

Dosyalar