Finite element analysis of bending and torsional loading of two different endodontic rotary files with different off-center geometric sections

dc.contributor.authorEskibaglar, Mehmet
dc.contributor.authorErdem, Serkan
dc.contributor.authorKaman, Mete Onur
dc.contributor.authorOcak, Mevlut Sinan
dc.date.accessioned2026-08-12T18:08:39Z
dc.date.issued2023
dc.departmentFırat Üniversitesi
dc.description.abstractThe aim of this study was to evaluate the performance of two endodontic rotary files with different off-center geometries in bending and torsional tests using finite element analysis (FEA). Trunatomy Prime (TRN-P) and WaveOne Gold Primary (WOG-P) files were used in the study. Geometric models for finite element analysis were obtained by micro-computed tomography scanning. FEA analyses were performed according to ISO 3630-1 specifications to evaluate the bending and torsional tests of the files. Trunatomy Prime showed higher flexibility than WOG-P when subjected to bending. In terms of torsional strength, WOG-P showed higher torsional strength than TRN-P. The geometry of a rotary instrument file influences the mechanical behavior (bending and torsion) of rotary files. The clinician should be aware of the behavioral differences in rotary files and use the appropriate file according to the clinical situation in addition to the manufacturer's instructions.
dc.identifier.doi10.1515/mt-2023-0070
dc.identifier.endpage1712
dc.identifier.issn0025-5300
dc.identifier.issn2195-8572
dc.identifier.issue11
dc.identifier.orcid0000-0002-3121-2116
dc.identifier.orcid0000-0002-2504-0135
dc.identifier.scopus2-s2.0-85169795668
dc.identifier.scopusqualityQ2
dc.identifier.startpage1707
dc.identifier.urihttps://doi.org/10.1515/mt-2023-0070
dc.identifier.urihttps://hdl.handle.net/11508/63156
dc.identifier.volume65
dc.identifier.wosWOS:001052642000001
dc.identifier.wosqualityQ1
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.language.isoen
dc.publisherWalter de Gruyter Gmbh
dc.relation.ispartofMaterials Testing
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/openAccess
dc.snmzKA_WoS_20260511
dc.subjectbending and torsional loading
dc.subjectclinical fatigue
dc.subjectcross section
dc.subjectendodontic devices
dc.subjectfinite element analysis
dc.titleFinite element analysis of bending and torsional loading of two different endodontic rotary files with different off-center geometric sections
dc.typeArticle

Dosyalar