Evaluation of 4 Point-of-Care Units for the Determination of Blood L-Lactate Concentration in Cattle

dc.contributor.authorKarapinar, T.
dc.contributor.authorKaynar, O.
dc.contributor.authorHayirli, A.
dc.contributor.authorKom, M.
dc.date.accessioned2026-08-12T17:48:04Z
dc.date.issued2013
dc.departmentFırat Üniversitesi
dc.description.abstractBackgroundDespite being used commonly in bovine medicine, information on reliability of point-of-care (POC) lactate meters is limited. ObjectiveTo determine the validity of 4 commercially available POC lactate meters in cattle. AnimalsCattle with various diseases (n=120). MethodsBlood samples collected from the jugular vein were processed immediately on POC lactate meters. Plasma l-lactate concentration was measured by the enzymatic-colorimetric method (ELISA). Data were subjected to Friedman's test for comparison, Passing-Bablok regression and Bland-Altman plot analyses for reliability, and receiver operating characteristics analysis for sensitivity (Se) and specificity (Sp). ResultsThe POC lactate meters were highly correlated with ELISA (r=0.98-0.99) despite disagreements among units. When regressed on ELISA, blood l-lactate concentrations generated from Accutrend Plus and Lactate Pro were linear up to 16.6 and 15.7mmol/L, respectively, whereas those generated from i-STAT and Lactate Scout were linear up to 19.5 and 19.7mmol/L, respectively. All POC lactate meters had a Se of 100% with Sp of 95.7-98.6% at a plasma l-lactate cut-off concentration of 4mmol/L. i-STAT had the best accuracy (99.0%) and precision (99.8%), the best linear fit (y=-0.13+1.04X) yielding the lowest bias (-6.6%) as well as the highest Se (100%) and Sp (98.6%). Conclusions and Clinical ImportanceDespite high correlation with the reference method, dilution is needed for Accutrend Plus/Lactate Pro and i-STAT/Lactate Scout if concentrations >15 and 20mmol/L, respectively. i-STAT provided the most accurate and precise results.
dc.description.sponsorshipScientific and Technological Research Council of Turkey (TUBITAK) project [111O045]
dc.description.sponsorshipThis research was supported by The Scientific and Technological Research Council of Turkey (TUBITAK) project no: 111O045. We thank Asst. Prof. Bulent Cavusoglu (Department of Electrical and Electronics Engineering, Ataturk University) for MATLAB usage.
dc.identifier.doi10.1111/jvim.12204
dc.identifier.endpage1603
dc.identifier.issn0891-6640
dc.identifier.issn1939-1676
dc.identifier.issue6
dc.identifier.pmid24118259
dc.identifier.scopus2-s2.0-84887615675
dc.identifier.scopusqualityQ1
dc.identifier.startpage1596
dc.identifier.urihttps://doi.org/10.1111/jvim.12204
dc.identifier.urihttps://hdl.handle.net/11508/61284
dc.identifier.volume27
dc.identifier.wosWOS:000326972600040
dc.identifier.wosqualityQ1
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherWiley-Blackwell
dc.relation.ispartofJournal of Veterinary Internal Medicine
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.snmzKA_WoS_20260511
dc.subjectAccuracy
dc.subjectBovine
dc.subjectPortable lactate meter
dc.subjectPrecision
dc.subjectSensitivity
dc.subjectSpecificity
dc.titleEvaluation of 4 Point-of-Care Units for the Determination of Blood L-Lactate Concentration in Cattle
dc.typeArticle