Evaluation of the Analytical and Diagnostic Performances of a Point-of-Care Device for Hematocrit Measurement in Calves

dc.contributor.authorTumer, Kenan C.
dc.contributor.authorBabacan, Sumeyye
dc.contributor.authorWittek, Thomas
dc.date.accessioned2026-08-12T17:11:17Z
dc.date.issued2025
dc.departmentFırat Üniversitesi
dc.description.abstractObjective To evaluate the performance of a point-of-care hemoglobin/hematocrit measurement device (POCHD), validated for human medicine, for measuring hematocrit in calves. Design Prospective study conducted between July 2022 and April 2023. Setting University teaching hospital. Animals One hundred thirty-three calves <1 month old, suffering from various diseases. Interventions The microhematocrit method (MHM) and a POCHD were used to measure hematocrit in venous blood samples. The MHM served as the reference method. The analytical performance of the POCHD was evaluated using Bland-Altman plots and calculating the intraassay coefficient of variation and observed total error (TEobs). Diagnostic performance was assessed by implementing receiver operating characteristic curve analysis and Cohen's kappa statistic. Measurements and Main Results Bland-Altman plots showed a mean bias of -0.03 L/L (-3.26%) (95% confidence interval: -0.04 to -0.03 [-4.02% to -2.51%]). The intraassay coefficient of variation for the POCHD was 2.4% for low hematocrit values and 0.7% for high hematocrit values. TEobs ranged from 3.2% to 32.9% (median: 18.1%). Receiver operating characteristic curve analysis at hematocrit thresholds of <0.22 L/L (22%) for anemia and >0.33 L/L (33%) for hemoconcentration showed optimized thresholds of <0.19 L/L (18.5%) and >0.33 L/L (33.2%), respectively, for the POC device to differentiate anemic from nonanemic calves, as well as calves with dehydration-related hemoconcentration from those without. At the thresholds of 0.22 L/L (22%) and 0.33 L/L (33%), Cohen's kappa statistic found agreements of 90.2% (kappa = 0.536) and 91% (kappa = 0.820), respectively, between the POCHD and the MHM. Conclusions Clinicians should be aware that the POCHD underestimates hematocrit values and exhibits high TEobs; however, when adjusted thresholds are applied, the POCHD shows good sensitivity and specificity for distinguishing anemia and dehydration-related hemoconcentration.
dc.identifier.doi10.1111/vec.70048
dc.identifier.endpage520
dc.identifier.issn1479-3261
dc.identifier.issn1476-4431
dc.identifier.issue5
dc.identifier.orcid0000-0002-2861-0236
dc.identifier.orcid0000-0001-8007-8315
dc.identifier.pmid41085138
dc.identifier.scopus2-s2.0-105018957601
dc.identifier.scopusqualityQ2
dc.identifier.startpage514
dc.identifier.urihttps://doi.org/10.1111/vec.70048
dc.identifier.urihttps://hdl.handle.net/11508/51091
dc.identifier.volume35
dc.identifier.wosWOS:001593239000001
dc.identifier.wosqualityQ3
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherWiley
dc.relation.ispartofJournal of Veterinary Emergency and Critical Care
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.snmzKA_WoS_20260511
dc.subjectanemia
dc.subjectcalves
dc.subjectdehydration
dc.subjectmethod comparison
dc.subjectvalidation
dc.titleEvaluation of the Analytical and Diagnostic Performances of a Point-of-Care Device for Hematocrit Measurement in Calves
dc.typeArticle

Dosyalar