1 fm/?Hz noise level low temperature Fabry-Perot atomic force/magnetic force microscope operating in 4-300 K temperature range

dc.contributor.authorKarc, Ozgur
dc.contributor.authorCelik, Umit
dc.contributor.authorOral, Ahmet
dc.date.accessioned2026-08-12T17:18:20Z
dc.date.issued2020
dc.departmentFırat Üniversitesi
dc.description.abstractIn this publication, we describe the design of a new fiber Fabry-Perot interferometer and employ this to a low temperature atomic force/magnetic force microscope (LT-AFM/MFM) operating in the 4-300 K temperature range. A multilayer dielectric mirror coated optical fiber is used to achieve unprecedented 1 fmnoise level, while the shot noise limit is 0.51 fm. The cavity length is adjustable, and the fiber can be brought within a very close proximity of the cantilever using a dedicated 2 mm stroke piezonanopositioner integrated on the piezotube scanner. The same nanopositioner also is used to park the fiber at a safe parking location during cantilever exchange. We demonstrate the performance of the LT-AFM/MFM by imaging the ITO thin film at 300 K, atomic steps on HOPG at 300 K, magnetic bits on the hard disk at 15 K, and the Abrikosov vortex lattice in the BSCCO(2122) single crystal at 4 K. Published under license by AIP Publishing.
dc.description.sponsorshipTUBITAK [111T802, 2212]
dc.description.sponsorshipWe thank TUBITAK for partial financial support (Grant No. 111T802). We are grateful to Professor K. Kadowaki of Tsukuba University for supplying the high quality single crystal BSSCO(2212) sample. We thank Dr. Ashvin S. Cherodian of NanoMagnetics Instruments for revising the manuscript.
dc.identifier.doi10.1063/1.5120007
dc.identifier.issn0034-6748
dc.identifier.issn1089-7623
dc.identifier.issue1
dc.identifier.orcid0000-0002-7759-6821
dc.identifier.orcid0000-0002-0704-8491
dc.identifier.pmid32012622
dc.identifier.scopus2-s2.0-85078223269
dc.identifier.scopusqualityQ2
dc.identifier.urihttps://doi.org/10.1063/1.5120007
dc.identifier.urihttps://hdl.handle.net/11508/53006
dc.identifier.volume91
dc.identifier.wosWOS:000519796100001
dc.identifier.wosqualityQ3
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherAip Publishing
dc.relation.ispartofReview of Scientific Instruments
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/openAccess
dc.snmzKA_WoS_20260511
dc.subjectInterferometer
dc.subjectCantilevers
dc.title1 fm/?Hz noise level low temperature Fabry-Perot atomic force/magnetic force microscope operating in 4-300 K temperature range
dc.typeArticle