Reliability of the i-STAT for the determination of blood electrolyte (K+, Na+, and CI-) concentrations in cattle
| dc.contributor.author | Yildirim, E. | |
| dc.contributor.author | Karapinar, T. | |
| dc.contributor.author | Hayirli, A. | |
| dc.date.accessioned | 2026-08-12T17:48:25Z | |
| dc.date.issued | 2015 | |
| dc.department | Fırat Üniversitesi | |
| dc.description.abstract | Background: Rapid determination of blood electrolyte concentrations can help determine electrolyte status and delivery of effective volume of electrolyte solutions in field conditions. Objective: To evaluate reliability of the i-STAT, a point-of-care (POC) device, in measuring blood K+, Na+, and CI- concentrations in cattle. Animals: Ninety-eight cattle with various diseases. Methods: In this prospective study, blood samples collected from the jugular vein were processed for determination of K+, Na+, and CI- concentrations in blood and plasma using the i-STAT and auto-analyzer (Cobas C501), respectively. Blood and plasma electrolyte data were subjected to student t-test for comparison, the concordance analysis for agreement, accuracy, and precision, the Passing-Bablok regression and the Bland-Altman plot for reliability, and receiver operating characteristics curves for sensitivity (Se) and specificity (Sp). Results: Plasma concentrations of K+ (4.39 versus 4.2 mmol/L; P <.0001) and CI- (100.30 versus 99.4 mmol/L; P <.04) were greater than their concentrations in blood. Plasma and blood Na+ concentrations were similar (136.95 versus 136.8 mmol/L). The i-STAT results were highly correlated with the Cobas C501 results (r = 0.970, 0.922, and 0.866 for K+, Na+, and CI-, respectively). Regression equations fitting blood (Y) and plasma (X) concentration did not deviate from the identity line for K+ (Y = - 0.10 + 0.98 9 X), Na+ (Y = X), and CI- (Y = 3.04 + 0.96 9 X). The mean bias (blood concentration -plasma concentration) was - 0.20 for K+ (P =.03), -0.16 for Na+ (P = .12), and -0.87 for CI- (P = .93). The i-STAT had 76-100% Se and 87.7-100% Sp for assessing electrolyte statuses. Conclusions and Clinical Importance: The i-STAT yielded results that were in agreement with the auto-analyzer, with negligible biases in measurement of plasma K+, Na+, and CI- concentrations. The i-STAT is a reliable POC device and can be used in field condition to assess electrolyte status in cattle. | |
| dc.description.sponsorship | Firat University Scientific Research Unity (FUBAP) [VF.13.02] | |
| dc.description.sponsorship | This research was supported by Firat University Scientific Research Unity (FUBAP) project #: VF.13.02. | |
| dc.identifier.doi | 10.1111/jvim.12526 | |
| dc.identifier.endpage | 394 | |
| dc.identifier.issn | 0891-6640 | |
| dc.identifier.issn | 1939-1676 | |
| dc.identifier.issue | 1 | |
| dc.identifier.pmid | 25619525 | |
| dc.identifier.scopus | 2-s2.0-84921483068 | |
| dc.identifier.scopusquality | Q1 | |
| dc.identifier.startpage | 388 | |
| dc.identifier.uri | https://doi.org/10.1111/jvim.12526 | |
| dc.identifier.uri | https://hdl.handle.net/11508/61419 | |
| dc.identifier.volume | 29 | |
| dc.identifier.wos | WOS:000348664000055 | |
| dc.identifier.wosquality | Q1 | |
| dc.indekslendigikaynak | Web of Science | |
| dc.indekslendigikaynak | Scopus | |
| dc.indekslendigikaynak | PubMed | |
| dc.language.iso | en | |
| dc.publisher | Wiley-Blackwell | |
| dc.relation.ispartof | Journal of Veterinary Internal Medicine | |
| dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | |
| dc.rights | info:eu-repo/semantics/openAccess | |
| dc.snmz | KA_WoS_20260511 | |
| dc.subject | Accuracy | |
| dc.subject | Bovine | |
| dc.subject | Point-of-care device | |
| dc.subject | Precision | |
| dc.subject | Sensitivity | |
| dc.subject | Specificity | |
| dc.title | Reliability of the i-STAT for the determination of blood electrolyte (K+, Na+, and CI-) concentrations in cattle | |
| dc.type | Article |